Parameter Fluoride Iodide Code HI4010 HI4110 FC301B HI4011 HI4111 Type solid-state; half-cell solid-state; combination solid-state; half-cell solid-state; half-cell solid-state; combination Measurement Range 1M to 1•10-⁶M Sat. to 0.02 mg/L (ppm) 1M to 1•10-⁷M 127000 to 0.01 mg/L (ppm) Optimum pH Range 5 to 8 5 to 8 5 to 8 2 to 13 2 to 13 Temperature Range 0 to 80°C 0 to 80°C 0 to 80°C 0 to 80°C 0 to 80°C Approximate Slope -56 -56 -56 -56 -56 Body O.D. 12 mm 12 mm 12 mm 12 mm 12 mm Insertion Length 120 mm 120 mm 120 mm 120 mm 120 mm Body Material epoxy PEI/epoxy epoxy epoxy PEI Cable 1 m coaxial 1 m coaxial 1 m coaxial 1 m coaxial 1 m coaxial Possible Applications determination of free fluoride in potable water, soft drinks, wine, plants, emulsified food products, plating and pickling acids determination of free iodide ions in emulsified food samples (iodized table salt), plants and for titration Connection BNC BNC BNC BNC BNC HI4010 · HI4110 · FC301B · HI4011 · HI4111 Ion Selective Electrodes Fluoride · Iodide 120 mm 012 mm 120 mm 012 mm 120 mm 012 mm 120 mm 012 mm 120 mm 012 mm 3 ISE 3.29 www.hannainst.com | sensors