EPA Compliant The HI83414 meets and exceeds the requirements of EPA and Standard Methods both for turbidity and colorimetric chlorine measurements. When in EPA mode all turbidity readings are rounded accordingly to meet reporting requirements. Backlit Graphic LCD Display A graphic LCD display provides an easy to understand, user-friendly interface. All messages are in plain text making them easy to read. Light blocking cuvette cover An a©xed, light-blocking cuvetter cover closes over the sample cell, reducing stray light from aecting any measurement readings. Four Measurement Modes The HI83414 features four measurement modes including ratio or non-ratio mode for turbidity, free chlorine, and total chlorine. In ratio mode the turbidity is 0.00 to 4000 NTU (Nephelometric Turbidity Units) while in the non-ratio mode the range is 0.00 to 40.0 NTU. The range for free or total chlorine measurements is 0.00 to 5.00 mg/L (ppm) range. Multiple Turbidity Units of Measure Turbidity can be displayed as nephelometric turbidity units (NTU), European Brewing Convention units (EBC) or Nephelos units. Multiple reading modes Normal measurement, continuous measurement, or signal averaging measurement are reading modes available CAL Check™ With the powerful CAL Check function, reliable performance of the chlorine colorimeter can be validated at any moment by using the exclusive HANNA ready-made, NIST traceable standards. All standards are supplied with a Certicate of Analysis (COA) for traceability. Calibration A two, three, four, or ve-point turbidity calibration can be performed by using the supplied (<0.1, 15, 100, 750, and 2000 NTU) standards. Calibration points can be modied if user-prepared standards are used. For free and total chlorine, the CAL Check standard can be used for calibration to 1.00 mg/L (ppm). AMCO AEPA-1 Primary Turbidity Standard The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are stable and reusable with a long shelf life. These standards are used for calibration and performance verication of the HI83414 turbidity meter. 12 Turbidity EPA meters 12.7 www.hannainst.com |
RkJQdWJsaXNoZXIy MTI3NTM=